Instructions for using DN40 spiral vortex flowmeter
Instructions for using DN40 spiral vortex flowmeter

Instructions for using DN40 spiral vortex flowmeter

Category:

Instrumentation/Flow meter/flowmeter

Model:

LUX-40

Brand:

Jiangsu Fenghui Instrument Co., Ltd

Retail Price

2,800.00USD


重量

kg

  • Product Description
  • Description :

      This product adopts a built-in combination to make the structure more compact. It can directly measure the flow rate, pressure, and temperature of fluids, and automatically track compensation and compression factor correction in real time.

      1、 Product features:

      No mechanical movable parts, not easily corroded, stable and reliable, long service life, long-term operation without special maintenance.

      The shell is available in stainless steel and aluminum alloy options.

      Built in pressure, temperature, and flow sensors, high safety performance, compact structure, and beautiful appearance; It can directly measure fluid pressure, temperature, and flow rate, and automatically track compensation and compression factor correction in real time.

      Adopting advanced computer processing chips, it has high integration, small size, good performance, and strong overall functionality. It adopts a circuit micro power consumption design, with low overall power consumption. It can be powered by an internal battery (3.6VDC) or an external power supply (24VDC).

      2、 Technical parameters:

      Measuring fluids: natural gas and other gases

      Caliber series: DN25-DN200

      Accuracy level: ± 1.00/, ± 1.50%

      Repeatability: less than one-third of the absolute value of the basic error limit

      Working power supply: 24-28VDC (two-wire system), external 12-28VDC (three wire system), dry battery

      Output signal: frequency pulse signal, 4-20mA, RS485 communication

      Electrical interface: M20X1.5

      Protection level: IP65

      Environmental temperature: -30 ℃~+55 ℃

      Relative humidity: 5% to 95%

      Atmospheric pressure: 86KPa~106Kpa

      Working conditions: Medium temperature: -20 ℃ --+70 ℃

      Nominal pressure: 1.6, 2.5, 4.0, 6.3 MPa

      Pipeline connection: flange connection

      3、 Attention:

      1. The directional signs can be installed in vertical, horizontal, or any inclined position.

      2. When the pipeline is long or close to the vibration source, supports should be installed upstream and downstream of the flowmeter to eliminate the impact of pipeline vibration.

      3. The installation location of the vortex flowmeter should have sufficient space for inspection and maintenance of the flowmeter, and should meet the environmental requirements of the flowmeter.

      4. Avoid interference from strong external magnetic fields.

      5. When installed and used outdoors, there should be a cover to avoid direct sunlight and rainwater erosion, which may affect the service life of the instrument.