Integrated LUGB Vortex Flow Meter LUGB-32 Remote Transmission Vortex Flow Meter
Category:
Instrumentation/Flow meter/flowmeter
Model:
LUGB-32
Brand:
Jiangsu Fenghui Instrument Co., Ltd
Retail Price
1,450.00USD
重量
kg
- Product Description
-
Description :
1、 Overview:
This product is widely used in the measurement of fluids in industries such as petroleum, chemical, metallurgy, and papermaking. The flowmeter has no movable parts, strong reliability, high accuracy, and long service life. It can accurately measure the instantaneous flow rate and cumulative flow rate of liquids over a wide flow range.
2、 Technical parameters:
Nominal diameter (mm): 25, 40, 50, 65, 80, 100, 125, 150, 200, 250, 300, 300 (insert type)
Nominal pressure (MPa): DN25-DN200 4.0 (>4.0 protocol supply), DN250-DN300 1.6 (>1.6 protocol supply)
Medium temperature (℃): Piezoelectric type: -40~260, -40~320; Capacitive: -40~300, -40~400, -40~450 (protocol ordering)
Body material: 1Cr18Ni9Ti, (supplied under other material agreements)
Allowable vibration acceleration: piezoelectric type: 0.2g, capacitive type: 1.0-2.0g
Accuracy: ± 1% R, ± 1.5% R, ± 1FS; Insertion: ± 2.5% R, ± 2.5% FS
Range: 1:6-1:30
Supply voltage: Sensor:+12V DC,+24V DC; Transmitter:+12V DC,+24V DC; Battery powered type: 3.6V battery
Output signal: Square wave pulse (excluding battery powered type): High level ≥ 5V, Low level ≤ 1V; Current: 4-20mA
Pressure loss coefficient: in accordance with JB/T9249 standard Cd ≤ 2.4
Explosion proof mark: Intrinsic safety type: Exd Ⅱ ia CT2-T5 Explosion proof type: Exd Ⅱ CT2-T5
Protection level: Regular IP65, Submersible IP68
Environmental conditions: temperature -20 ℃~55 ℃, relative humidity 5%~90%, atmospheric pressure 86~106kPa
Applicable media: gas, liquid, steam

3、 Selection:
1. Diameter of pipeline
2. The name of the tested medium (steam should indicate whether it is saturated steam or superheated steam)
3. Working pressure of the tested medium
4. Working temperature of the tested medium
5. The working flow rate of the tested medium
4、 Select installation location:
1. The installation position should be selected as much as possible in a location without pipeline vibration or with low vibration, and the vibration acceleration should not exceed 2g. If the vibration is large, vibration reduction measures need to be taken;
2. There must be sufficient straight pipe sections upstream and downstream of the flowmeter;
3. The maintenance valve is installed upstream of the sensor, and the flow control valve is installed downstream of the sensor;
4. Try to choose a location that is easy to install and maintain;
5. Choose a location with a dry environment;
6. Sensors can be installed on horizontal or vertical pipelines. But when installed on a vertical pipeline, the medium must flow from bottom to top.
7. Sensors should be installed indoors as much as possible, and attention should be paid to waterproofing when installed outdoors. Cables outside the amplifier should be bent into a U-shape.

AfterSalesService :
Key words:- Integrated LUGB Vortex Flow Meter