Fenghui Instrument LUXB-100 Spiral Vortex Flow Meter
Category:
Instrumentation/Flow meter/flowmeter
Model:
LUX-100
Brand:
Jiangsu Fenghui Instrument Co., Ltd
Retail Price
3,950.00USD
重量
kg
- Product Description
-
Description :
Fenghui Instrument LUXB-100 Spiral Vortex Flow Meter
This product has the advantages of strong functionality, wide flow range, simple operation and maintenance, and easy installation and use. Its main technical indicators have reached the advanced level of similar foreign products as a new type of gas flow meter.
1、 Characteristics:
1. No mechanical movable parts, not easily corroded, stable and reliable, long service life, long-term operation without special maintenance;
2. Using 16 bit computer chips, high integration, small size, good performance, and strong overall functionality;
3. The intelligent flowmeter integrates a flow probe, microprocessor, pressure, and temperature sensors, and adopts a built-in combination to make the structure more compact. It can directly measure the flow rate, pressure, and temperature of the fluid, and automatically track compensation and compression factor correction in real time;
4. Adopting dual detection technology can effectively improve the strength of detection signals and suppress interference caused by pipeline vibration;
5. Intelligent seismic technology effectively suppresses interference signals caused by vibration and pressure fluctuations;
6. Adopting a Chinese character dot matrix display screen with multiple digits, the reading is intuitive and convenient. It can directly display the volume flow rate under working conditions, the volume flow rate under standard conditions, the total amount, as well as parameters such as medium pressure and temperature.

Fenghui Instrument LUXB-100 Spiral Vortex Flow Meter
2、 Technical performance:
execution standard
The flowmeter shall comply with JJG 198-94 "Verification Regulations for Velocity Flowmeters".
Natural cyclone inlet vortex flowmeter flange connection type
Environmental conditions for use
Environmental temperature: -30 ℃~55 ℃
Relative humidity: 5% to 95%
Atmospheric pressure: 86KPa~106KPa
working conditions
Medium temperature range: -20 ℃ to 70 ℃
Nominal pressure: 1.6, 2.5, 4.0, 6.3 MPa
3、 Installation precautions:
1. The intelligent vortex flowmeter can be installed in vertical, horizontal, or any inclined position according to the flow direction indicator.
2. When the pipeline is long or close to the vibration source, supports should be installed upstream and downstream of the flowmeter to eliminate the impact of pipeline vibration.
3. The installation location of the vortex flowmeter should have sufficient space for inspection and maintenance of the flowmeter, and should meet the environmental requirements of the flowmeter.
4. Avoid interference from strong external magnetic fields.
5. When installed and used outdoors, there should be a cover to avoid direct sunlight and rainwater erosion, which may affect the service life of the instrument.
6. During pipeline pressure testing, attention should be paid to the pressure measurement range of the pressure sensor equipped with the intelligent vortex flowmeter to avoid damage to the pressure sensor due to overpressure.
7. Attention should be paid to the influence of installation stress. The upstream and downstream pipelines of the flowmeter should be coaxial, otherwise shear stress will be generated. The installation location of the flowmeter should consider the thickness of the sealing gasket or install an elastic expansion joint on the downstream side.
8. Before installing the intelligent vortex flowmeter, welding slag and other debris in the pipeline should be removed first.

AfterSalesService :
Key words:- LUXB Spiral Vortex Flow Meter