Description :
High precision LDG-50 electromagnetic flowmeter
This product has good wear resistance, excellent elasticity, high tensile strength, and a working temperature of less than 80 ℃. It is suitable for measuring general water, sewage, mud, and slurry, and is resistant to corrosion from low concentration acid-base media.
1、 Characteristics:
1. Provide display interfaces in both Chinese and English;
2. It can directly display the instantaneous flow rate, cumulative flow rate, percentage flow rate, and flow velocity in both forward and reverse directions according to the engineering unit;
3. There are 6 types of flow units to choose from (L/m, L/h, m3/m, m3/h, GPM, GPH);
4. 4~20mADC load 0~750, scale flow rate (upper limit of range) can be set arbitrarily;
5. By setting the pulse coefficient, each pulse can represent a certain volume flow rate, or output a 0-2kHz frequency signal proportional to the flow rate;
6. Provide traffic upper and lower limit reporting function;
7. RS485 interface, bidirectional communication. The working parameters of the flowmeter can be set and modified through the upper computer;
8. When the fluid is stationary, the zero point of the flowmeter can be automatically calibrated through the buttons on the panel;
9. Damping adjustment can be set within a range of 1 to 99 seconds;
10. Power supply voltage: 220VAC/24VDC type;
11. Environmental temperature: -20~60 ℃;
12. Protection level: IP65;
13. Explosion proof grade: ExdibIIBT4.

High precision LDG-50 electromagnetic flowmeter
2、 Technical parameters:
Instrument accuracy: pipeline type 0.5 level, 1.0 level; Insertion type 2.5 level
Measurement medium: various liquid and liquid-solid two-phase fluids with conductivity greater than 5 μ S/cm.
Flow velocity range: 0.2-8m/s
Work pressure: 1.6MPa
Environmental temperature: -40 ℃ to+50 ℃
Medium temperature: PTFE lining ≤ 180 ℃
Rubber material lining ≤ 65 ℃
Explosion proof mark: Exmibd Ⅱ BT4
Explosion proof certificate number: GYB01349
External magnetic interference: ≤ 400A/m
Shell protection: Integrated type: IP65
3、 Attention:
1. Determination of applicable traffic range. The selection of instrument diameter is determined based on the flow range. For electromagnetic flow meters, they are generally selected in situations where the flow signal is strong and the instrument can work stably for a long time. We suggest that you choose your instrument caliber based on the flow rate flow comparison table (see table below) and refer to the actual operating conditions. We recommend a flow rate range of 1-3m/s, which is beneficial for the instrument to be in a suitable state.
2. Selection of lining and electrodes. In addition to determining the diameter and flow range, electromagnetic flowmeters also need to choose relevant conditions such as sensor lining and matching electrodes to improve their economic applicability. The design unit can choose different linings, electrodes, etc. according to the different media of different users
3. Pressure and usage situation. In addition to determining the above conditions, the electromagnetic flowmeter should also improve the selection of operating pressure, installation form, shell protection level, explosion-proof requirements, output configuration, and other usage conditions and conditions. The integrated type is suitable for indoor environments with good performance, while the split type is suitable for situations where the sensor is used in harsh environments, with large pipeline vibrations and high temperature of the measured medium. Please choose according to the instrument model markings.

AfterSalesService :
Key words:- LDG electromagnetic flowmeter