COD online analyzer, real-time monitoring of chemical oxygen demand, environmental certification online COD detector
Category:
Environmental Protection/Analytical instruments/Water quality online monitoring system
Model:
SZ-COD
Brand:
Jingcheng Instrument
model:
SZ-COD
Working Voltage:
AC230±10%V
measurement accuracy:
± 10% or ± 5mg/L (whichever is greater)
structural style :
cabinet-type
Temperature control range:
Suggested temperature: 15-25 ℃
Product Type:
Online instruments
Processing & Customization:
No
power:
50—60Hz,5A
weight:
fifty
Scope of Application:
0—500mg/L 50—2000mg/L 100—5000mg/L
Place of Origin:
Qingdao
manufacturer:
Qingdao Jingcheng
Retail Price
18,500.00USD
重量
kg
- Product Description
-
model SZ-COD
Working Voltage AC230±10%V
measurement accuracy ± 10% or ± 5mg/L (whichever is greater)
structural style cabinet-type
Temperature control range Suggested temperature: 15-25 ℃
Product Type Online instruments
Processing & Customization No
power 50—60Hz,5A
weight fifty
Scope of Application 0—500mg/L 50—2000mg/L 100—5000mg/L
Place of Origin Qingdao
manufacturer Qingdao Jingcheng
Description :

COD online analyzer
Eco-Certification
1、 SZ-COD COD online analyzer product overview:
Chemical Oxygen Demand (COD) refers to the mass concentration of oxygen consumed by strong oxidants under certain conditions when oxidizing organic and inorganic reducing substances in water samples. COD is also an important indicator reflecting the degree of pollution of water bodies by organic and inorganic reducing substances.
This analyzer is capable of long-term automatic continuous operation according to on-site settings, and is widely applicable to industrial pollution source discharge wastewater, industrial process wastewater, industrial wastewater treatment plant wastewater, municipal wastewater treatment plant wastewater, and other occasions. According to the complexity of the on-site testing conditions, corresponding preprocessing systems can be selected to ensure the reliability of the testing process, accurate testing results, and fully meet the on-site requirements of different occasions.
24-hour real-time monitoring
2、 SZ-COD COD online analyzer productPrinciple:
The water sample, digestion solution, liusuan silver solution (liusuan silver as a catalyst can effectively oxidize linear fatty compounds), and concentrated liusuan mixture are heated to 175 ℃. After the organic matter in the solution is oxidized by dichromate ions, the color will change. The analyzer detects this color change and converts it into COD value output. The amount of dichromate ions consumed is equivalent to the amount of oxidizable organic matter, which is the COD value.
Chemical Oxygen Demand Testing
3、 Technical parameters of SZ-COD COD online analyzer:
Specification Name
Technical specifications and parameters
measurement range
(switchable)
0—500mg/L
50—2000mg/L
100—5000mg/L
measurement method
Determination of Chemical Oxygen Demand - Dichromate Method
detection limit
1mg/L
resolution
zero point one
accuracy
± 10% or ± 5mg/L (whichever is greater)
repetitiveness
5%
zero drift
±5mg/L
Span Drift
±10%
measurement period
The measurement cycle for Z is 20 minutes, and the digestion time can be set between 5-60 minutes
sampling period
Interval, hourly, and trigger modes can be adjusted
calibration cycle
Automatic calibration 1-99 days adjustable
maintenance cycle
Over 1 month, approximately 1 hour per session
Self checking protection
Abnormal or power outage data not lost
After abnormal reset or power outage, residual liquid will be automatically eliminated and work will resume
data storage
1 year of data storage
output interface
1 RS232 channel; Route 1 4-20mA
working environment
Suggested temperature: 15-25 ℃
Operating Humidity
≤ 90% (non condensing)
Power Supply and Power
AC230±10%V;50—60Hz,5A
size
1500*550*450mm
Online water quality detector
AfterSalesService :
Key words:- COD online analyzer