Sampling flow rate 5-20L, flow accuracy below 2.5%, high flow explosion-proof sampling pump
Category:
Instrumentation/Analytical instruments/sampler
Model:
FCC-1500D
Brand:
Jingcheng Instrument
model:
FCC-1500D
type:
air sampler
Installation method:
portable
display mode:
LED display
Sampling flow range:
2-25
power supply voltage:
220 or 12
ambient temperature:
10~35℃
ambient humidity:
≤95%
Processing & Customization:
yes
Sampling flow accuracy:
≤±2.5%
sampling time:
four hundred and eighty
noise:
one
weight:
three
Place of Origin:
Qingdao, Shandong
manufacturer:
Qingdao Jingcheng Instrument Co., Ltd
Retail Price
1,500.00USD
重量
kg
- Product Description
-
model FCC-1500D
type air sampler
Installation method portable
display mode LED display
Sampling flow range 2-25
power supply voltage 220 or 12
ambient temperature 10~35℃
ambient humidity ≤95%
Processing & Customization yes
Sampling flow accuracy ≤±2.5%
sampling time four hundred and eighty
noise one
weight three
Place of Origin Qingdao, Shandong
manufacturer Qingdao Jingcheng Instrument Co., Ltd
Description :
FCC-1500D Explosion proof Atmospheric Sampler, High Flow Explosion proof Sampling Pump, Occupational Health Explosion proof Atmospheric Sampling Pump 20-500ml/min
1、 Sampling flow rate 5-20L, flow accuracy below 2.5%. Introduction to high flow explosion-proof sampling pump:
The sampling flow rate is 5-20L, and the flow accuracy is less than 2.5%. The high flow explosion-proof sampling pump is widely used as a new instrument for environmental health monitoring, occupational health monitoring, industrial and mining enterprises, scientific research workplaces, and other collection workplaces that come into actual contact with toxic and harmful substances. The instrument adopts microcomputer chip control technology and is equipped with a high-energy cadmium nickel rechargeable battery. It can run continuously for up to 10 hours on a single charge. It is easy to operate, has high timing accuracy, is sturdy and durable, and is convenient and fast for timing selection. The countdown time can be freely selected from 0-99 hours and 59 minutes, and the power supply will be automatically turned off when the machine is stopped. It is currently an ideal sampler among similar instruments in China.
2、 The FCC-1500D explosion-proof atmospheric sampler mainly refers to the following technical parameters:
Sampling flow range (optional): 0-1L/min, 0.1-1.5L/min, 0.1-3L/min, 1-5L/min (new standard)
Flow error: ≤ 5%
Timing range: freely set within 99 hours and 59 minutes
Continuous working time:>8 hours
Timing error: ≤ 1 ‰
Negative pressure: ≥ 35000pa
Working temperature: 0 ℃~40 ℃
Storage temperature: -40 ℃ to+60 ℃
Environmental atmospheric pressure: 80KPa~110Kpa
Relative humidity: ≤ 95%
Noise: ≤ 60dB
The control button allows for pausing sampling during breaks, meals, or other situations
Anti backflow system: It can prevent liquids and particles from being sucked back into the instrument, thereby extending its service life
Explosion proof form: Intrinsically safe Exib Ⅱ AT3
Power supply: 10VDC, 220V 50Hz AC
Weight: Approximately 0.5kg
AfterSalesService :
Key words:- Explosion proof sampling pump
More Products