The online total phosphorus automatic analysis and detection instrument has complete manufacturer certificates and excellent equipment stability
Category:
Instrumentation/Analytical instruments/Water quality analyzer
Model:
LB-2120
Brand:
Qingdao Lubo Jianye Environmental Protection Techn
model:
LB-2120
measurement range:
0-10 mg/L
measurement accuracy:
Not exceeding ± 10%
power supply voltage:
AC220±10% V,50±10% Hz,5A
resolution:
Not exceeding ± 10%
weight:
60KG
Place of Origin:
Qingdao
manufacturer:
Qingdao Road Expo
Retail Price
99,999.00USD
重量
kg
- Product Description
-
model LB-2120
measurement range 0-10 mg/L
measurement accuracy Not exceeding ± 10%
power supply voltage AC220±10% V,50±10% Hz,5A
resolution Not exceeding ± 10%
weight 60KG
Place of Origin Qingdao
manufacturer Qingdao Road Expo
Description :
Directly supplied by Qingdao Lubo manufacturer, the manufacturer has been researching and developing for many years, and the equipment stability is excellent. The manufacturer provides on-site service, installation training, and complete certificates. Welcome to inquire by phone
Online total phosphorus automatic analysis and detection instrument
Product Introduction
The online total phosphorus automatic analyzer heats a mixture of water sample, catalyst solution, and strong oxidant digestion solution to 120 ℃. Polyphosphate and other phosphorus containing compounds in the water sample are digested and oxidized by the strong oxidant under acidic conditions of high temperature and high pressure to generate phosphate ions. In the presence of a catalyst, phosphate ions form a colored complex in a strong acid solution containing molybdate. The analyzer detects the color change and converts it into total phosphorus value output. The amount of colored complex generated is equivalent to the total phosphorus content.
Widely used in various fields such as sewage treatment plants, washing plants, chemical wastewater discharge outlets, industrial production process water, and industrial and municipal wastewater treatment.

Technical Specifications
Method basis
Phosphorus molybdenum blue spectrophotometric detection
measurement range
0-10 mg/L total phosphorus (graded as 0.1-2.0 mg/L); 0.2-5.0mg/L; 0.5-10.0mg/L; )10-500 mg/L total phosphorus (to be expanded)
accuracy
Not exceeding ± 10% or not exceeding ± 0.2mg/L
repetitiveness
Not exceeding ± 5% or not exceeding ± 0.2mg/L
measurement period
The minimum measurement cycle is 30 minutes, and according to the on-site environment, the color development time can be modified arbitrarily within 5-120 minutes
sampling period
Time interval (adjustable from 10 to 9999 minutes) and hourly measurement mode
calibration cycle
Adjustable at any interval and time from 1 to 99 days
maintenance cycle
Usually once a month, about 30 minutes each time
output
2-channel RS-232; Route 1 4-20mA
Environmental Requirements
For indoor environments with adjustable temperature, it is recommended to have a temperature range of+5-28 ℃; Humidity ≤ 90% (non condensing)
power supply
AC220±10% V,50±10% Hz,5A
AfterSalesService :
Key words:- Online total phosphorus automatic analysis and detection instrument