Application of comprehensive atmospheric sampler in environmental monitoring
Application of comprehensive atmospheric sampler in environmental monitoring

Application of comprehensive atmospheric sampler in environmental monitoring

Category:

Instrumentation/Environmental testing instruments/Atmospheric sampler

Model:

LB-2031A

Brand:

Qingdao Road Expo

brand:

Qingdao Road Expo

model:

LB-2031A

type:

Dual channel atmospheric sampler

flow range:

(60~140)L/min

Continuous working hours:

eight

Operating Temperature:

(-55~125)℃

power supply:

AC 220V(±10%)

Processing & Customization:

yes

Overall dimensions:

398 × 193 × 288mm (length × width × height)

weight:

About 13

Place of Origin:

Qingdao

manufacturer:

Qingdao Road Expo

Retail Price

7,580.00USD


重量

kg

  • Product Description
  • brand

    Qingdao Road Expo

    model

    LB-2031A

    type

    Dual channel atmospheric sampler

    flow range

    (60~140)L/min

    Continuous working hours

    eight

    Operating Temperature

    (-55~125)℃

    power supply

    AC 220V(±10%)

    Processing & Customization

    yes

    Overall dimensions

    398 × 193 × 288mm (length × width × height)

    weight

    About 13

    Place of Origin

    Qingdao

    manufacturer

    Qingdao Road Expo

    Description :

    Qingdao Lubo manufacturer produces environmental monitoring instruments with numerous devices. Here are a few examples on the internet

    List of Equipment Procurement for Third Party Testing Unit Laboratory: Equipment Names: Constant Temperature and Humidity Weighing System, Atmospheric Sampler, X - γ Dose Rate Meter, Personal Radiation Dosimeter, Alpha - β Surface Pollution Analyzer, Multi channel Gamma Spectrometer, Neutron Peripheral Dose Equivalent Meter, Electric Blast Drying Oven, COD Constant Temperature Heater, Portable Gas Dust and Smoke Sampling Instrument Calibration Device, Intelligent Electronic Soap Film Flow Meter, Digital Anemometer, Precision Temperature and Humidity Meter, Precision Noise Spectrum Analyzer, Ultra Pure Water Machine, Portable Turbidity Meter, Ultrasonic Cleaner, Handwheel Stainless Steel Pressure Sterilizer, Flow Meter, Smoke Blackness Tester, Water Sulfide Blower, Intelligent Integrated Distillation Instrument, Atomic Absorption Spectrophotometer, Oscillator, Electronic Balance, Visible Light Spectrophotometer, Perforation Meter Extraction instrument, biochemical incubator, portable conductivity meter.

    LB-2031A Integrated Atmospheric Sampler

      Product Introduction: This instrument uses the filter membrane weighing method to capture total suspended particulate matter (TSP), inhalable micro particles (PM10), or fine particulate matter (PM2.5) in the ambient air. Collect various gaseous substances in the ambient atmosphere and indoor air using solution absorption method. It can be widely used in environmental monitoring, health and epidemic prevention, labor protection, scientific research institutions and other fields for sample collection of gaseous substances and particulate matter.

    Technical Specifications

    Main parameters

      

    parameter range

      

    resolution

      

    accuracy

      

    Dust sampling flow rate

      

    (60~140)L/min , Working point flow rate of 100 L/min

      

    0.1L/min

      

    ≤±2%

      

    Atmospheric sampling flow rate

      

    (0.1-1.0) L/min (expandable)

      

    0.1L/min

      

    ≤±2%

      

    sampling time

      

    Set freely within 0-99H59 minutes

      

    1min

      

    20min ≤±1s

      

    Pre measurement temperature

      

    (-55~125)℃

      

    0.1℃

      

    ≤±2.0℃

      

    Pre measurement pressure

      

    (-20~0)kPa

      

    0.01kPa

      

    ≤±2.5%

      

    atmospheric pressure

      

    (60~130)kPa

      

    0.01kPa

      

    ≤±500Pa

      

    Constant temperature control

      

    (15~30)℃

      

    0.1℃

      

    ≤±2.0℃

      

    Traffic repeatability

      

    ≤ 2%

      

    Traffic stability

      

    ≤ 5%

      

    load capacity

      

    At a flow rate of 100 L/min, it can overcome a resistance of 9kPa

      

    piece

    grain

    thing

    pick

    sample

    head

      

    Cut

    cut

    special

    conscript

      

    TSP sampling head

      

    Da50=(100±1)μm δg =1.5 ±0.1

      

    PM10 sampling head

      

    Da50=(10±0.5)μm δg =1.5±0.1

      

    PM2.5 sampling head

      

    Da50=(2.5±0.2)μm δg =1.2±0.1

      

    Entrance speed

      

    0.3 m/s

      

    Effective filter membrane diameter

      

    Φ80mm

      

    connector

      

    M27×1.5

      

    Allowable deviation of air inlet width

      

    ≤±2.0%

      

    Overall dimensions

      

    398 × 193 × 288mm (length × width × height)

      

    Sampler noise

      

    ≤62dB(A)

      

    working power supply

      

    AC 220V (± 10%), 50Hz, built-in lithium battery (13.6AH)

      

    AfterSalesService :

    Key words:
    • Integrated atmospheric sampler