COD online automatic monitoring instrument for water source data monitoring
Category:
Instrumentation/Analytical instruments/Water quality analyzer
Model:
CODcr
Brand:
Qingdao Lubo Jianye Environmental Protection Techn
model:
CODcr
type:
Rapid water quality analyzer
Installation method:
online
Processing & Customization:
yes
measurement range:
10-5000 mg/L COD
measurement accuracy:
Not exceeding ± 10%
power supply voltage:
AC220±10% V,50±10% Hz,5A
resolution:
Not exceeding ± 10%
weight:
About 40
Place of Origin:
Qingdao
manufacturer:
Qingdao Road Expo
Retail Price
16,500.00USD
重量
kg
- Product Description
-
model CODcr
type Rapid water quality analyzer
Installation method online
Processing & Customization yes
measurement range 10-5000 mg/L COD
measurement accuracy Not exceeding ± 10%
power supply voltage AC220±10% V,50±10% Hz,5A
resolution Not exceeding ± 10%
weight About 40
Place of Origin Qingdao
manufacturer Qingdao Road Expo
Description :
Factory supplied COD online monitoring instrument, chemical oxygen demand CODcr online water quality automatic analysis and detection instrument
1. Overview of CODcr Water Quality Online Automatic Monitoring Instrument
CODcr water quality online automatic monitoring instrument is currently mainly used in surface water, sewage, tap water, rainwater, rivers, farmland, enterprise production, environmental inspection and other places. The instrument can automatically complete data monitoring of on-site water sources, and transmit data to various required places such as mobile phones, computers, and the internet. The instrument has accurate and fast results, comes with an environmental certificate, automatic continuous monitoring, and achieves data networking function.
2. Method basis for CODcr water quality online automatic monitoring instrument
Determination of Chemical Oxygen Demand for Water Quality - Dichromate Method HJ 354-2019
Technical Requirements for Chemical Oxygen Demand CODCr Water Quality Automatic Monitoring Instrument HJ/T 377-2007
Online monitoring system for water pollution sources (CODcr)
Brief Introduction to the Principle of CODcr Water Quality Online Automatic Monitoring Instrument:
Take a quantitative water sample, add silver solution and concentrated mixture, and heat it to 165 ℃ in a digestion tank and maintain it for a period of time. During this period, chromium ions act as oxidants and are reduced from VI valence to III valence, changing color. The absorbance corresponds to the content of organic compounds in the sample, and the instrument directly displays the COD of the sample through colorimetric conversion. The main interfering substance of this method is chloride, which is automatically added before digestion

Technical Specifications
Method basis
Determination of Chemical Oxygen Demand - Dichromate Method
measurement range
10-5000 mg/L COD
accuracy
When ≥ 100mg/L, not exceeding ± 10%; When<100mg/L, not exceeding ± 8mg/L
repetitiveness
When ≥ 100mg/L, not exceeding ± 10%; When it is less than 100mg/L, it should not exceed ± 6mg/L.
measurement period
The minimum measurement cycle is 20 minutes, and according to the actual water sample, the digestion time can be modified arbitrarily within 5-120 minutes
sampling period
Time interval (adjustable from 10 to 9999 minutes) and hourly measurement mode
calibration cycle
Adjustable at any interval and time from 1 to 99 days
maintenance cycle
Usually once a month, about 30 minutes each time
output
RS-232, 4-20mA (optional)
Environmental Requirements
For indoor environments with adjustable temperature, it is recommended to have a temperature range of+5-28 ℃; Humidity ≤ 90% (non condensing)
power supply
AC220±10% V,50±10% Hz,5A