Repair of abnormal function of Yuyue oxygen concentrator: Beijing whole district appointment for on-site service
Category:
Business Services/Maintenance and installation services/Electrical maintenance and installation
Model:
complete series
Brand:
fish leaping
Retail Price
9,000,000.00USD
重量
kg
- Product Description
-
Description :
Yuyue oxygen concentrator maintenance appointment for on-site maintenance in Beijing area: 010-8963 6276
1、 Abnormal product functionality issues
Insufficient oxygen output or anaerobic output
Check if the water level in the humidification bottle is between the minimum Min and maximum Max markings, as a low water level can cause oxygen to not output properly.
Confirm if the filter screen is clogged. It is recommended to clean the filter cotton every 3 months and replace the filter element with a new one every six months.
Check if the placement environment of the oxygen concentrator is well ventilated, and leave a heat dissipation space of at least 30cm around to avoid high temperatures affecting the performance of the molecular sieve.
Equipment alarm prompts low oxygen concentration alarm
Immediately suspend use, check if the air intake is obstructed by foreign objects, clean and restart the device.
If the alarm continues, it is highly likely that the performance of the molecular sieve has decreased. It is necessary to contact the maintenance service of Yuyue oxygen concentrator for professional testing.
2、 Physical fault handling
Abnormal noise or vibration
Confirm if the machine is placed steadily and adjust the bottom adjustable foot pad to maintain a horizontal position.
Check if the fixing screws of the compressor are loose. It is recommended to perform a deep maintenance once a year for long-term use.
Display screen malfunction
After disconnecting the power and allowing it to stand for 10 minutes, restart the machine for testing. If there is still no display, it is mostly a motherboard fault and requires professional maintenance or return to the factory for repair.
3、 Daily maintenance suggestions
After the equipment has been running for 100 hours, ultraviolet disinfection can be carried out, and 75% medical alcohol can be used to wipe the humidification bottle.
Before long-term shutdown, it is recommended to run the machine without load for 2 hours, remove the moisture inside the machine, and then store it. The humidity of the storage environment should be controlled within 60%.
It is recommended to perform a professional oxygen concentration test and accuracy calibration every 5000 hours of machine operation.