Description :
4、 Configuration:
1. Host: One set (including vacuum pump, multifunctional controller)
2. Level 6 impactor: one unit
3. Tripod: one unit
4. Charger: one
5. Special accessories: one set
6. Operation manual: one copy
7. Warranty card: one copy
8. Certificate of conformity: one copy
9. Aluminum alloy suitcase: one


1、 BY-400 Air Microbial Sampler Execution Standard:
GBT 18204.3-2013 "Methods for hygiene inspection of public places (airborne microorganisms)"
JJF1826-2020 Calibration Specification for Air Microbial Samplers
2、 Introduction to BY-400 Air Microbial Sampler:
The BY-400 intelligent six level sieve impact air microbial sampler can automatically set the environmental air microbial sampling for 6 weeks. It is a product designed by our factory, with two sampling methods: single period sampling and multi period sampling. The sampling flow rate is large, the stability is good, and the power supply adopts AC/DC dual-use type, which is convenient to use. It has a small size and light weight, and can be widely used for monitoring in hygiene and epidemic prevention, biological cleanliness, pharmaceuticals, fermentation industry and other environments, as well as for sampling and research of air microorganisms in relevant research and teaching departments. It provides a basis for evaluating the harm of air environmental microbial pollution and its prevention and control measures.
3、 Technical parameters of BY-400 air microbial sampler:
The instrument adopts a multifunctional microcontroller control module, a large screen LCD display, Chinese menu, human-machine dialogue mode, and Chinese prompts when human intervention is required. This model is developed by our factory, and the instrument automatically controls the sampling flow rate. It uses a multifunctional controller to control constant current sampling and automatically compensates for flow changes caused by voltage and resistance changes. (Electronic flowmeter), with high sampling accuracy and stable flow rate.
Sampling principle: Principle of six level impact method (Anderson)
Flow range: 28.3L/min
Flow error: ≤± 5%
Traffic repeatability: ≤ 2%
Timing error: ≤± 0.1%
Measurement range: Capture rate: ≥ 98%
Extraction negative pressure: ≥ 2000pa
Collection device: using standard culture dishes
Can be automatically set arbitrarily: ① Sampling flow rate
② Single time sampling time
③ Multi time sampling time and sampling flow rate
④ Multi period sampling interval time
Intelligent timer timing range: 1 minute to 6 weeks, automatically set sampling time; Accuracy: 0.01%
Working voltage: AC/DC dual-use type, built-in rechargeable battery
Noise: ≤ 50 dB
Power: ≤ 45W
Capture particle range:
Grade:>7.0 μ m aperture 1.18mm
Second level: 4.7 μ m-7.0 μ m aperture 0.91mm
Third level: 3.3 μ m-4.7 μ m aperture 0.71mm
Fourth level: 2.1 μ m-3.3 μ m aperture 0.53mm
Level 5: 1.1 μ m-2.1 μ m aperture 0.34mm
Level 6: 0.65 μ m-1.1 μ m aperture 0.25mm