Integrated atmospheric sampler with built-in lithium battery
Integrated atmospheric sampler with built-in lithium battery

Integrated atmospheric sampler with built-in lithium battery

Category:

Instrumentation/Environmental testing instruments/Atmospheric sampler

Model:

LB-2030

Brand:

Qingdao Road Expo

brand:

Qingdao Road Expo

model:

LB-2030

type:

Dual channel atmospheric sampler

flow range:

60~140L/min

Continuous working hours:

eight

Operating Temperature:

See details

power supply:

AC 220V (± 10%), 50Hz or built-in lithium battery

Processing & Customization:

yes

Overall dimensions:

256 × 270 × 200mm (length × width × height)

weight:

About 5

Place of Origin:

Qingdao

manufacturer:

Qingdao Road Expo

Retail Price

8,900.00USD


重量

kg

  • Product Description
  • brand

    Qingdao Road Expo

    model

    LB-2030

    type

    Dual channel atmospheric sampler

    flow range

    60~140L/min

    Continuous working hours

    eight

    Operating Temperature

    See details

    power supply

    AC 220V (± 10%), 50Hz or built-in lithium battery

    Processing & Customization

    yes

    Overall dimensions

    256 × 270 × 200mm (length × width × height)

    weight

    About 5

    Place of Origin

    Qingdao

    manufacturer

    Qingdao Road Expo

    Description :

      Manufacturer contact person: Fu Yajian, Qingdao Lubo manufacturer specializes in environmental testing instruments. With over ten years of experience and excellent technology, Qingdao Lubo is worthy of your trust

      LB-2030 Integrated Atmospheric Sampler (with built-in lithium battery)

      

      Product Introduction

      This instrument uses the filter membrane weighing method to capture total suspended particulate matter (TSP), inhalable particulate matter (PM10), or fine particulate matter (PM2.5) in the ambient air. Collect various gaseous substances in the ambient atmosphere and indoor air using solution absorption method; It can be widely used in environmental monitoring, health and epidemic prevention, labor protection, scientific research institutions and other fields for sample collection of gaseous substances and particulate matter.

      execution standard

      JJG 956-2013 Atmospheric Sampler

      JJG 943-2011 "Total Suspended Particle Sampler"

      HJ/T 374-2007 Technical Requirements and Testing Methods for Total Suspended Particle Samplers

      HJ/T 375-2007 "Technical Requirements and Testing Methods for Environmental Air Samplers"

      HJ/T 93-2013 "Technical Requirements and Detection Methods for PM10 and PM2.5 Samplers in Environmental Air"

      HJ 618-2013 "Determination of PM10 and PM2.5 in ambient air by gravimetric method"

      Technical Specifications

      Main parameters

      

      parameter range

      

      resolution

      

      accuracy

      

      Dust sampling flow rate

      

      (60~140)L/min , Working point flow rate of 100 L/min

      

      0.1L/min

      

      ≤±2%

      

      Atmospheric sampling flow rate

      

      (0.1-1.0) L/min (expandable)

      

      0.1L/min

      

      ≤±2%

      

      sampling time

      

      Set freely within 0-99H59 minutes

      

      1min

      

      20min ≤±1s

      

      Pre measurement temperature

      

      (-55~125)℃

      

      0.1℃

      

      ≤±2.0℃

      

      Pre measurement pressure

      

      (-20~0)kPa

      

      0.01kPa

      

      ≤±2.5%

      

      atmospheric pressure

      

      (60~130)kPa

      

      0.01kPa

      

      ≤±500Pa

      

      Traffic repeatability

      

      ≤ 2%

      

      Traffic stability

      

      ≤ 5%

      

      load capacity

      

      At a flow rate of 100L/min, it can overcome a resistance of 9kPa

      

      piece

      grain

      thing

      pick

      sample

      head

      

      Cut

      cut

      special

      conscript

      

      TSP sampling head

      

      Da50=(100±1)μm δg =1.5 ±0.1

      

      PM10 sampling head

      

      Da50=(10±0.5)μm δg =1.5±0.1

      

      PM2.5 sampling head

      

      Da50=(2.5±0.2)μm δg =1.2±0.1

      

      Entrance speed

      

      0.3 m/s

      

      Effective filter membrane diameter

      

      Φ80mm

      

      connector

      

      M27×1.5

      

      Allowable deviation of air inlet width

      

      ≤±2.0%

      

      Overall dimensions

      

      256 × 270 × 200mm (length × width × height)

      

      Sampler noise

      

      ≤62dB(A)

      

      working power supply

      

      AC 220V (± 10%), 50Hz or built-in lithium battery

      

    AfterSalesService :

    Key words:
    • Integrated atmospheric sampler