High strength acid salt index CODMn online automatic monitoring instrument water quality online analyzer
Category:
Instrumentation/Analytical instruments/Water quality analyzer
Model:
RC-601( xrp)
Brand:
Thermal Holding Environmental Protection
model:
RC-601( xrp)
type:
Rapid water quality analyzer
Installation method:
online
Processing & Customization:
yes
measurement range:
0-20mg/L ; 0-50mg/L ; Customizable range
measurement accuracy:
Detection limit of ≤± 10%: 0.5mg/L
power supply voltage:
two hundred and twenty
resolution:
one
weight:
twenty
Place of Origin:
Qingdao
manufacturer:
Thermal Holding Environmental Protection
Retail Price
25,000.00USD
重量
kg
- Product Description
-
model RC-601( xrp)
type Rapid water quality analyzer
Installation method online
Processing & Customization yes
measurement range 0-20mg/L ; 0-50mg/L ; Customizable range
measurement accuracy Detection limit of ≤± 10%: 0.5mg/L
power supply voltage two hundred and twenty
resolution one
weight twenty
Place of Origin Qingdao
manufacturer Thermal Holding Environmental Protection
Description :
CODMn online automatic monitoring instrument and water quality online analyzerTechnical Specifications
The permanganate index is a commonly used indicator to reflect the pollution of organic and inorganic oxidizable substances in water bodies. It is defined as the equivalent amount of oxygen calculated from the consumption of certain organic and inorganic reducing substances in an oxidized water sample under certain conditions. The index cannot be used as an indicator of theoretical oxygen demand or total organic matter content, because under specified conditions, many organic matter can only be partially oxidized, and volatile organic matter is not included in the measured values. The chemical oxygen demand measured using a solution as an oxidant, formerly known as manganese method chemical oxygen demand. In China's new environmental water quality standards, the value has been renamed as permanganate index, and only the value measured by the acidic method is referred to as chemical oxygen demand. The International Organization for Standardization (ISO) recommended method is limited to the determination of surface water, drinking water, and domestic sewage, and is not applicable to industrial wastewater. [3]
Measurement method: Oxidation titration method
Measurement range: 0-20mg/L; 0-50mg/L ; Customizable range
Indication error: ≤± 10% Detection limit: 0.5mg/L
Repeatability: ≤± 5% Waste liquid volume: 6mL/time
Memory effect: ≤± 3% Actual water sample comparison: ≤± 10%
Constant temperature time: 360 seconds Voltage stability: ≤± 5%
Measurement cycle: 30 minutes Calibration cycle: any time
Maintenance cycle: ≥ 720h/time Maintenance workload:<2 hours/month
Reagent consumption: 3 months/500ml Display output: Equipped with a 10.1-inch color LCD touch screen
Color rendering temperature: The recommended temperature is 160 ℃, which can be set according to the actual water sample situation
Data export: Measurement values can be imported into a USB drive through a USB port and saved
Signal output: RS485/RS232/USB interface/standard 2-channel 4-20mA output/standard 2-channel switch input/output
Environmental requirements: Temperature controlled indoor environment, recommended temperature (5-28 ℃), humidity ≤ 90% (no condensation)
Power supply and output: (220V ± 22) V/AC, (50 ± 0.5Hz), 5A, 150W
Instrument size: Upper cabinet 601 * 450 * 301mm; Upper cabinet 701 * 450 * 301mm
CODMn Online Automatic Monitoring Instrument
1、 Principle
Add a known amount of sum to the sample, heat it at high temperature for a certain period of time, and remove the sample
Some organic and inorganic reducing substances in the product are oxidized, and excess sodium oxalate is added after the reaction to reduce the residue
Yu, drip excess sodium oxalate back into the standard solution and calculate the sample
Medium potassium permanganate index value



2、 Product Features
The new photoelectric quantitative technology is not affected by interference such as high chromaticity, suspended solids, and bubbles, with high reliability and accuracy, and strong anti-interference ability.
High detection accuracy, low measurement limit, and small long-term drift of the instrument.
It has one click self-test function and self diagnostic function.
Adopting a single channel highly integrated valve group, maintenance and cleaning are simple and fast.
24/7 networking function, monitoring instrument operation status anytime, anywhere.
It has alarm function, quality control function, and reverse control function.
It has automatic cleaning and calibration functions.
Capable of whole point measurement, interval measurement, and external control measurement.
Automatically work after power failure and recovery.
Automatic prompt and reset function for faults and missing reagents (samples).
L has fault recording function.
L can save historical data for more than 3 years.
Ensure separate discharge of cleaning solution and waste liquid
III
AfterSalesService :
Key words:- Online water quality analyzer